Evaluation of the Quality of Medico-technical Equipment Sterilization in National University Hospital of Cotonou in Benin in 2013

All published articles of this journal are available on ScienceDirect.

RESEARCH ARTICLE

Evaluation of the Quality of Medico-technical Equipment Sterilization in National University Hospital of Cotonou in Benin in 2013

The Open Public Health Journal 31 Aug 2016 RESEARCH ARTICLE DOI: 10.2174/1874944501609010053

Abstract

Backgrounds:

In low income countries, hospital-acquired infections continue to develop in hospitalized patients, and may also affect medical staff. Medico-technical equipment sterilization is critical for prevention and safety care of nosocomial infections.

Objective:

To assess the quality of medico-technical equipment sterilization at the National University Hospital of Cotonou in 2013.

Method:

This cross-sectional and evaluative study was conducted at the National University Hospital of Cotonou from 10th June to 04th July 2013. A sample of 51 health workers involved in the of medico-technical equipment sterilization system, two (02) administrative authorities, the responsible of National Committee for the Fight against nosocomial infections in the hospital, 41 sterilized instruments and compresses were assessed in the study. Health workers were observed in their work environment before undergoing an individual interview as well as the administrative authorities and the Responsible of the National Committee for the Fight against nosocomial infections. Sterilized instruments are analyzed in microbiology laboratory.

Results:

More than half of the participants were male (52.9%). The average age of respondents was 41 ± 7.5 years. The sterilization unit of the hospital was managed by common surgical department of the hospital and its mission was to provide sterile medico-technical equipment. The sterilization unit did not meet the standard architecture of sterilization environment. Equipment sterilization procedure did not meet standards of quality assurance. There was no preventive maintenance procedure for autoclave and poupinel that were used for sterilization of instruments. No indoor cleaning and air sterilization of the service of sterilization were planned. However, equipment sterilization supplies were available, and 13.72% of workers surveyed were well-skilled. Microbiological tests showed that 48.8% of sterilized medical equipment was contaminated by Staphylococcus aureus, Pseudomonas aeruginosa and Enterobacter cloacae.

Conclusion:

The quality of instrument sterilization system in the HKM National University hospital of Cotonou was poor. Sterilized equipment was contaminated by pathogens. Medical equipment sterilization process needs improvement to prevent hospital-acquired infections.

Keywords: Benin, Hospital, Medico-technical equipment, Microorganism, Nosocomial infection, Sterilization.

INTRODUCTION

Hospital acquired infections remain an important public health concern. Preventing Nosocomial Infections (NI) is then a major objective for hospitals that aimed at providing safe health care to patients. The use of disposable equipment is the best way to ensure patient comfort and security. However, low income country health system could not acquire disposable format for certain devices due to the cost. Then some instruments need to be sterilized for re-use [1]. Sterilization is a set of methods that aims at eliminating all living organisms of any kind or form whatsoever carried on a perfectly cleaned object [2]. Sterilization helps reducing the number of germs (egg: 1,000,000 to 1) by six logarithms fold, while disinfection aims at reducing germs by 5 logarithms fold [3]. Hence, the importance of cleaning and disinfection before sterilization is critical as it is essential for ensuring that medical and surgical instruments do not transmit infectious pathogens to patients. It is recognized that any equipment could be well sterilized only if it is well cleaned [4, 5]. According to the manual on the Individualized Nursing (SFIS), sterilization is an activity related to health care, which can also be seen as an indirect care [6]. Sterilization helps preventing health care related infections and its quality reflects dysfunctions in hospital institutions.

Sterilization also aims at protecting everyone in the hospital: patients, medical staff and other hospital users. It is therefore essential for proper sterilization to establish a sterilization system based on standard references. According to the guidelines of Good Pharmacy Practices Hospital (BPPH) in France, proper treatment of medico-technical equipments requires human resources, adequate material and financial resources, trained personnel and a “quality” approach in order to prevent and reduce any risk for patients, professionals and the environment [7]. There are relatively few published assessments of equipment sterilization procedures from low income countries in general and Benin in particular. In Benin, a study reported an overall prevalence of nosocomial infections in 19.1% in hospitalized patients [8]. Despite this high frequency of nosocomial infection, little is known about the quality of instrument sterilization system. In Benin, there are no regulations along the health care quality in hospitals including instrument sterilization, despite guidelines for disinfection and sterilization in healthcare facilities being available since 2008 [9]. The objective of the study was to assess the quality of the sterilization system of medical equipment in the National University Hospital of Cotonou in Benin in 2013.

MATERIALS AND METHODOLOGY

Settings

The study was carried out at the National University hospital called CNHU-HKM of Cotonou.

The CNHU-HKM is a hospital that covers a land of 10 hectares with a capacity of 617 beds. In the last five years, the average number of patients seen in this hospital per year is 192600 [10]. Various missions are assigned to this National Hospital, including mainly:

  • Investigate and provide appropriate care for complex medical cases;
  • Train medical and related sector students; and
  • Conduct scientific research in conjunction with schools and health training institutes.

Main sterilization methods used in the hospital are “moist heat method” and “dry heat method”. The first uses autoclaves which are machines for sterilizing things such as surgical instruments and hospital equipments. They are a bit like giant pressure cookers that use the steam under pressure to kill off bacteria, spores and germs resistant to boiling water and powerful detergents. Because water is heated in a closed container, temperatures above 100°C can be reached. The last method uses “poupinel” for sterilization by hot air (dry heat) at elevated temperature. This method is convenient for metal, heat resistant glass, and vaseline, but is not convenient for linen or gauze swabs [2].

Study Design

This was a cross-sectional and evaluative study conducted to assess the quality of the sterilization system of medical equipments in the National University Hospital of Cotonou in Benin in 2013.

Study Population

The study population consisted of primary and secondary targets. The primary targets included sterilized medico-technical equipments (instruments and compresses) and workers in CNHU-HKM. The secondary targets included the responsible of the committee of fight against nosocomial infections (CLIN) and administrative authorities of the hospital (the Director and manager of human resources). Eligible participants who had not given their consent were excluded from the study.

Sampling

The sampling method used was non probabilistic for the workers who were selected by convenience, while administrative authorities were identified using rational choice.

The sampling method was probabilistic for sterilized medico-technical equipment (instrument and compresses) using the simple random technique.

The sample of materials study included:

  • Fifty-one (51) health workers involved in the sterilization process, which were selected using convenience technique;
  • Two (02) administrative authorities and the president of CLIN 01, which were selected according to the technique of rational choice; and
  • Forty-one (41) medical and technical equipments and swabs, which were collected using simple random selection.

Study Variables

The main variable of the study is the quality of sterilization that has two modalities:

  • First modality: Good (no germ found),
  • Second Modality: Poor (presence of germs),

The main variable is the quality of the sterilization system of medico-technical equipment.

The components that influence the main variable are “the inputs”, “the process” and “the result” according to the model “quality improvement” of Donabédian [11].

The component “inputs” included:

  • The infrastructure’s environment, the existence of the dirty areas, the appearance of walls, ceilings and surfaces, the frequency of equipment cleaning, and the existence of instruction that prohibited entrance to anyone outside the service;
  • The material resources concerning their availability, quality, functionality and effective use;
  • Equipment (containers, trolleys, metal cabinets);
  • Consumables: detergents, disinfectants, brushes, gloves, filters, corner strips, paper, canvas bags or packaging;
  • Guideline documents for sterilization process including material traceability records, filing procedures, maintenance contracts, supplier catalogs, administrative records and databases; and
  • Human resources namely health workers involved in the process of medico-technical equipment sterilization and the activities/actions relative to compliance with hospital hygiene standard precautions (hand hygiene, personal hygiene, wearing of protective barriers, immunization and prevention of accidental exposure to blood).

The component “Process” referred to the application of technical and administrative procedures in the sterilization unit.

  • Administrative procedures included:
    • the existence and implementation of a work plan;
    • the existence of a flowchart;
    • the existence of a job description;
    • the existence of a workers’ training plan;
    • the existence of a monitoring - assessment plan of service activities;
    • the supervision of staff; and
    • the staff motivation mechanism.
  • Technical Procedures included:
    • written and validated sterilization procedures in force in the unit;
    • existence and implementation of a policy of procurement and maintenance of autoclave and poupinel used for sterilization;
    • promotion of hospital hygiene;
    • controls the water and air into the sterilization department;
    • achievement of physical and chemical controls;
    • carrying out bacteriological controls; and
    • achievement of visual ant tactile controls.

The component “result” referred to the quality of the medico-technical instrument sterilization that was assessed by the presence or absence of germs on the sterilized material.

Microbiological Analysis

The bacteriological analysis of medico-technical instruments and compresses was performed in surgery services and other medico-technical services of CNHU-HKM just after the opening of the instrument or compresses containers. Sterilized equipments were randomly selected. Sterile swabs were used to pass and/or scouring the equipments just after opening the container in the surgical operating unit or in wound repair service. The collection was introduced in incubated broth (37°C) oven for 24 hours. After 24 hours, the broth was seeded in required solid area (Agar fresh Blood, Chapman, Mueller Hinton, Eosin Methylene Blue) for 24 hours at 37°C. Positive cultures were subjected to additional tests (Gram test and other) and then the API-20E gallery for Entero-bacteria and others (Gram negative bacteria), the Coagulasse, Dnase and grouping tests for Cocci Gram-positive were carried out (Staphylococcus and Streptococcus).

Definition and Modalities of Variables

Operational definition and modalities of variables are described in Table 1.

Table 1.
Definition and modalities of variables.
Variables Definition Modalities
Qualification of workers Worker who have received basic training in a medical or paramedical training school Yes/No
Staff trained in sterilization procedures Existence of a sterilization training certificate or a diploma after trainings Yes/No
Availability of workers Three (03) or more workers are present every day for sterilisation activities Yes/No
Knowledge of sterilization procedures Worker knows in the required order of the steps of sterilization Yes/No
Observance of steps of sterilization In practice, the agent follows the normal order of the sterilization process Yes/No
Performance The worker knows and respects the different steps of sterilization Yes/No
Choice of sterilization method The agent knows how to choose the method of sterilization depending on the nature of the material, packaging, destination and level of treatment required Yes/No
Choice of sterilization cycle The worker knows how to choose sterilization cycle based on the level of treatment required Yes/No
Physical sterilization control The responsible of the unit controls the time, temperature and pressure according to the sterilizing cycle Yes/No
Physico-chemical sterilization control The responsible of the unit checks the strips bend at the end of the sterilization process Yes/No
Bacteriological control
of sterilization
Presence or absence of germ on the sterilized equipment Yes/No
Water control for sterilization Existence of report on the quality of water used during sterilization process Yes/No
Air control for sterilization Existence of the air monitoring report during sterilization process Yes/No
Access to sterilization unit Access to service is limited Yes/No

Measurement of Study Variables

The variables were operationalized as follows. A major criterion for assessing the quality of material was chosen: the presence of germs on the medical-technical equipment after sterilization.

Each practice was assessed as follows: when the practice meets the standards, it was marked “YES”, and when the practice does not meet the standards it was marked “NO” (YES = 1 and NO = 0).

Each component was assessed “Good” if all the items that go into its composition were “YES” and there were no germs on the sterilized equipment.

Even if only one practice was listed “NO”, the component was rated “Bad.”

The quality of sterilization was “Good” if all the components were “Good”, otherwise it was rated “Bad.”

Even if all the components were rated “Good” and a germ was found on sterilized medical and technical equipment, the quality of material was assessed “Bad.”

The quality of the sterilization was “good” if all the components or the quality of medico-technical sterilization system were assessed “Good” and no germ was found on the sterilized medico-technical equipment.

A major criterion for assessing the quality of material was chosen: the presence of germs on the medical-technical equipment after sterilization.

For each component, it was assessed as follows: when the practice meets the standards required (YES), score 1 was assigned and when the practice does not meet the standards (NO), the score 0 was assigned. Each component was appreciated “Good” if all the elements that compose it met the respective required standards. Otherwise the component was appreciated “Bad”.

The quality of the sterilization was “good” if all the components or the quality of medico-technical sterilization system were appreciated “Good” and no germ was found on sterilized medico-technical equipment as shown in Table 2.

Table 2.
Scores used for sterilization system evaluation.
Components Number of variables accessed Score Component Result Quality of
sterilisation system
Environmental factors 6 Yes = 1
No = 0
R1 Good or Bad
Human and material resources 4 Yes = 1
No = 0
R2
Technical and administrative procedures 5 Yes = 1
No = 0
R3
Organizational factors 12 Yes = 1
No = 0
R4

Data Collection Procedures

Data were collected using various techniques:

  • Direct observation was used for the appreciation of the unit of sterilization and the performance of workers regarding compliance with the steps of sterilization;
  • The document review was done to check traceability of the sterilized equipment and procedures used;
  • Structured individual interviews were used to collect information about administrative authorities of the hospital and the responsible of CLIN;
  • Laboratory analyses were performed seeking for germs on sterilized equipment.

Ethical Considerations

The authorization to conduct the study was obtained from the Director of HKM CNHU. The objectives of the study were explained to the participants. Voluntary informed consent was obtained from each participating worker before starting the interview. This consent claims that the participants were not at risk by refusing to participate in the survey or stopping their collaboration during the study any way prior to commencing the interview. The confidentiality and anonymity of the information collected were respected.

Data Analysis

Data were analyzed using Epi Info version 3.5.3. Quantitative variables were expressed as mean and standard deviation, and categorical variables were expressed as percentages.

RESULTS

Socio-demographic Characteristics, Knowledge and Observance of Steps of Sterilization in Participants

Socio-demographic characteristics, knowledge and observance of steps of sterilization in participants are described in Table 3 as follows:

Table 3.
Participants’ socio-demographic characteristics, knowledge and observance of steps of sterilization at HKM CNHU in Benin, 2013.
Variables Frequencies (n=51) Percentage (%)
Sex
    Male 27 52.9
    Female 24 47.1
Occupation
    Nurse/Mid wife 26 51
    Nurse assistant 15 29.4
    Laboratory skilled agents 06 11.8
    Trainees 04 7.8
Training in sterilization
    Yes 28 54.9
    No 23 45.1
Knowledge of steps sterilization
    Yes 12 23.5
    No 39 76.5
Observance of steps of sterilization
    Yes 13 25.5
    No 38 74.5
The average age of respondents was 41 ± 7.5 years. Table shows that 45.1% of workers have not received training in the field of sterilization.

Distribution of Workers According to Their Performance Regarding Medico-technical Equipment Sterilization at HKM CNHU in Benin in 2013

Fig. (1) below summarizes proportion of workers with adequate or inadequate performance regarding medico-technical equipment sterilization at HKM CNHU in Benin in 2013.

Fig. (1). Distribution of workers according to performance in medico-technical equipment sterilization at HKM CNHU of Cotonou in Benin, 2013.

Fig. (1) shows that 86.3% of workers involved in sterilization had inadequate (low) performance, while 13.7% exhibited adequate (high) performance.

Structural and Organizational Environment of Medico-technical Equipment Sterilization at HKM CNHU

Structural and organizational environment of medico-technical equipment sterilization at HKM CNHU is described in Table 4.

Table 4.
Structural and organizational environment of medico-technical equipment sterilization at HKM CNHU and European standards.
Variables Compliance with standards
Yes No
Structural aspects
Walls X
Ceilings X
Surfaces /tables X
Local air conditioner X
Cleaning according to the procedures X
Limited access to sterilization unit X
Human and material Resources
Autoclaves X
Consumables X
Individual protection X
Availability of human resources X
Control of quality
Existence of water quality monitoring report X
Existence of the air quality monitoring report X
Physical controls (pressure, temperature, duration) X
Chemical controls (Bowie and Dick test, shift indicator of passage, integrators) X
Bacteriological controls X
Organizational concerns
Material traceability document X
Recording of sterilization cycles X
Equipment maintenance contract X
Existence sterilization procedures X
Standardisation of sterilization procedures X
Centralization of the sterilization process X
Existence of local cleaning procedures X
Existence of organizational chart X
Existence of a work plan X
Existence of a training course plan X
Existence of a monitoring and evaluation plan X
Existence of supervision reports X

Among the items studied, few elements complied with European standards. Among the six items regarding respect to the structural aspect, only 1 (17%) conformed to European standards. Among the 12 items considered for organizational aspect, 2 (17%) complied with European standards. According to the human and material resources, 3 out of 4 (75%) items complied with European standards. Overall, 6 (22.2%) items out of 27 conformed to European standards.

Classification of Germs Isolated from Sterilized Medico-technical Equipment

The classification of germs isolated from sterilized medico-technical equipment in CNHU-HKM in 2013 is shown in Fig. (2), Tables 5 and 6.

Table 5.
Bacteria isolated from moist heat (autoclave) or dry heat (poupinel) sterilized medico-technical equipments in different surgeon operating rooms in HKM CNHU of Cotonou, Benin, 2013.
Surgery department* Sterilized materiel Mode of sterilization Bacteria found Bacteria counting/ Number of colonies counted
Emergencies service Instrument Poupinel Bacteria Gram(+) ≥ 500
Compresses Autoclave Coagulase-negative Staphylococcus ≥ 500
Clinic of gynecology and obstetrics Instrument Poupinel None 0
Compresses Poupinel Pathogen Gram(+) 200-500
Ophthalmology Instrument Autoclave None 0
Compresses Autoclave Coagulase-negative Staphylococcus 200-500
Urology room 1 Instrument Autoclave None 0
Compresses Autoclave None 0
Visceral surgery room 2 Instrument Autoclave Staphylococcus
Saprophyticus
≥ 500
Instrument Autoclave Pathogens Gram(+) ≥ 500
Pediatric surgery room 3 Instrument Autoclave None 0
Instrument Autoclave None 0
Traumatology room 4 Instrument Autoclave None 0
Instrument Autoclave None 0
Surgery Room 7 Instrument Autoclave Coagulase-negative Staphylococcus ≥ 500
Compresses Autoclave None 0
*Surgery department refers to the source of the instruments and compresses analyzed.
Table 6.
Bacteria isolated from moist heat (autoclave) or dry heat (poupinel) sterilized medico-technical equipments in different wound care services CNHU-HKM.
Wound care Service* Material analyzed Sterilization mode Bacteria identified Bacteria counting/ Number of colonies counted
Pediatric surgeon Instrument Poupinel None 0
Compresses Poupinel None 0
Stomatology Instrument Poupinel None 0
Compresses Poupinel None 0
Traumatology Instrument Poupinel None 0
Compresses Poupinel Gram negative Cocco bacillus ≥ 500
Clinic of gynecology and obstetrics Instrument Poupinel None 0
Compresses Poupinel Gram positive bacteria 200-500
Neonatology Compresses Poupinel Enterobacter cloacae 200-500
External surgeon Instrument Poupinel Staphylococcus aureus ≥ 500
Compresses Autoclave None 0
Visceral surgeon Instrument Poupinel Not identified Gram positive bacteria ≥ 500
Compresses Poupinel Not identified Gram positive bacteria ≥ 500
Urology Instrument Poupinel Not identified Gram positive bacteria ≥ 500
Compresses Poupinel Not identified Gram positive bacteria ≥ 500
Large burned room Instrument Poupinel Not identified Gram negative bacteria 200-500
Compresses Autoclave None 0
* Wound care Service refers to the source of the instruments and compresses analyzed.

Isolated bacteria Gram (-) were 51.2%, while isolated bacteria gram (+) were 48.8%.

Bacteriological culture of the 41 samples (medical-technical equipment and compresses) made in the various surgical blocks and other medical-technical services CNHU-HKM, just after the opening of the container showed that 20 (48.8%) of 41 materials were contaminated by bacteria. Among the identified bacteria, 50% Gram (-) bacteria included Enterobacter cloacae, Pseudomonas aeruginosa and other bacteria Gram (-) bacteria were not identified; whereas 50% Gram (+) bacteria included mostly Staphylococcus aureus, Staphylococcus saprophyticus and other bacillus.

The medico -technical equipment and compresses, sterilized in CNHU-HKM and used in operating theaters are mostly contaminated by coagulase-negative staphylococci.

The presence of pathogens, such as Staphylococcus aureus and Enterobacter cloacae, on instruments and compresses discredits the quality of the sterilization system of medical-technical equipment in various medical-technical services in the hospital.

Fig. (2). Percentage of germs isolated from sterilized medico-technical equipment in HKM CNHU of Cotonou, Benin, 2013.

Antibiotic Resistance Profile of Bacteria Isolated from Sterilized Medico-technical Material

Antibiotic resistance profile of bacteria isolated from sterilized medico-technical material is summarized in Table 7 below.

Table 7.
Antibiotic resistance profile of bacteria isolated from steam and poupinel sterilized medico-technical material in various CNHU-HKM services, Cotonou, Benin 2013.
Services Bacteria identified Number of Antibiotics tested
(N)
Efficacy Antibiotic
(%)
Mid-efficacy antibiotics
(%)
Non-efficacy antibiotics
(%)
Emergency service Coagulase negative Staphylococcus 30 83.3 0 16.6
Neonatology Enterobacter cloacae 25 24 8 68
External surgeon Staphylococcus aureus 30 86.6 0 13.3
Ophthalmology. Coagulase negative Staphylococcus 30 66.6 3.3 30
Cardiology Pseudomonas Aeruginosa 23 21.7 13 65.2
Large burned service Not identified Gram negative bacteria 22 86.3 0 36.4
Surgeon room 7 Coagulase negative Staphylococcus 26 34.6 3.8 61.5
Visceral surgeon room 2 Staphylococcus saprophyticus 28 82.1 3.6 14.3

Most germs isolated on sterilized medico-technical material in CNHU HKM-2013 showed a drug resistance for tested antibiotics.

DISCUSSION

The study explored the quality of medico-technical material sterilization in National University Hospital of Cotonou through the description of the structural and organizational environment of the sterilization unit and bacteriological analysis of sterilized equipment. We found that there was no-one responsible for the quality of medical equipment sterilization process at HKM-CNHU in contrast to the situation reported by Dubaele in France in 2000. According to Dubaele, the pharmacist, with the support of the management staff of the institution, implements the “quality system” for all sterilization operations. The pharmacist is the quality manager sterilization in the hospital [12, 13]. According to standards, sterilization should be centralized with objective to implement and respect the sterilization procedures for reusable medical devices under the responsibility of the head of sterilization. This was not the case in our study. At HKM-CNHU, workers in sterilization unit have never been supervised, evaluated or motivated. None of them had the opportunity to undergo training in medico-technical sterilization process despite the progress in science and technology. The material available (autoclave) was used without preventive maintenance. Subsequently, only one choice of sterilization cycle was possible since the autoclave was blocked due to lack of maintenance. This situation did not allow assessment of compliance with the steps of sterilization. This study reported a low performance rate of 13.723% in workers, which is far from 60% reported by Viayinon in 2003 [14].

In HKM-CNHU of Cotonou, nurses or midwives who were trained in preparing equipments for surgical care were responsible for the preparation of boxes (packaging), loading and unloading of the autoclave or poupinel. The step of packaging the instruments is obscured by lack of means to accomplish it. Workers without basic paramedic training or briefing of hospital hygiene were sometimes recruited for a short duration for cleaning, rinsing and drying of soiled material. The need of training of workers at unit of sterilization is critical, as well as the need of a new modern autoclave and better organization of work in the unit.

The compliance level of structural and organizational environment of the sterilization unit in HKM-CNHU compared to the European standards was low (22.2%). Indeed, sterilization by dry heat was still the most frequently used, while Poupinel method of sterilization is prohibited in Europe [15]. The circular 2001/138 of 14 March 2001 stipulated that this process is inefficient regarding Transmissible Unconventional Agents (NCTA) called “prions” or abnormal infectious protein, a general term given to agents that cause neurodegenerative disorders.

The profession of sterilization was not recognized in Benin. It was the same in France, but the first official recognition of the sterilizing agent profession was effective in February 2011. The certification of specialization of sterilization in hospital is issued by the Academic Centre for Continuing Education (CAFOC) of Toulouse [16].

Biological controls on sterilization units could be done in two ways using small tubes containing growing environment and bacterial spores in one hand and strips impregnated with bacterial spores for evaluation after autoclaving in the other hand.

Biological tests are not performed at CNHU-HKM due to lack of adequate equipments and their scarcity in the market.

It explained why we adopted the method which consisted of just swab sterile equipments at the opening of the container in the surgical operating room or in the dressing room with sterile swabs. The absence of sterilization standardized procedures leads to the fact that, in the hospital, each service disinfects, cleans and sterilizes the material in its own way.

For example, in stomatology department, packs of instruments are sterilized in an autoclave and benchtop instruments in poupinel for a period of two hours. This protocol established by the supervisor of the service was strictly respected by the workers. During checking, no germs were identified on compresses and sterilized instruments in the service.

Staphylococcus aureus were found on instruments sterilized by poupinel in the main unit of the hospital sterilization and on those being used in the surgery department [17]. The study allowed us to identify Pseudomonas aeruginosa on the heat-sensitive equipment disinfected and unsterilized in the cardiology department. A similar situation was reported by other studies [18, 19]. It also raises the issue of sterilization of heat-sensitive equipment in HKM- CNHU. Nowadays, recommended sterilization methods are gas or ionizing radiation. The method we used for biological control allowed identification of only germs, while the biological tests recommended but not available in Benin would allow assessing the effectiveness of the sterilization process. This method was used in the high school province in Namur in Belgium by Michau in the context of surgery specialization in pediatrics [20].

The Bowie and Dick test allows us to appreciate the smooth operation of the autoclave. It should be conducted every morning, while stopped for at least four hours or after any repair work on the autoclave. This was not implemented in CNHU-HKM in 2013.

Furthermore, it has been observed that the physical, chemical and biological tests have never been done in the sterilization unit of HKM-CNHU, despite knowing that this practice is an ongoing process to provide a guarantee of optimal quality care to patients [21].

CONCLUSION

This study highlighted the inadequacy of the architectural design of the sterilization unit of CNHU-HKM. The level of compliance of structural and organizational environment of the sterilization unit in HMK-CNHU was low as compared to European standards. The level of performance of the workers involved in the sterilization process was inadequate. Germs were identified on equipment marked as “sterile”. These findings demonstrated the poor quality of the sterilization system of medical-technical equipments at HKM CNHU of Cotonou in 2013.

Improvement of the quality of the sterilization system at HKM CNHU is critically needed in order to prevent hospitalized acquired infections.

CONFLICT OF INTEREST

The authors confirm that this article content has no conflict of interest.

ACKNOWLEDGEMENTS

The authors acknowledge the participants who have contributed to the study.

REFERENCES

1
Bretot Rihouey G. La stérilisation. Available from: http://medecine-pharmacie.univ-rouen.fr/servlet/com.univ.collaboratif.utils.Lecture Fichiergw?ID_FICHIER=10980 2007. [Accessed: 05/03/2016]
2
d’Hygiène CS. Recommandations en matière de stérilisation. Bruxelles: CSH 2006, no. 7848.
3
Rutala WA, Weber DJ. Disinfection and sterilization in health care facilities: what clinicians need to know. Clin Infect Dis 2004; 39(5): 702-9.
4
Haxhe JJ, Zumofen M. Hygiène hospitalière du matériel, 2016. Available from: http://www.md.ucl.ac.be/didac/hosp/cours/HH4.htm [Accessed: 05/03/2016]
5
Guimarães AC, Donalisio MR, Santiago TH, Freire JB. Mortality associated with nosocomial infection, occurring in a general hospital of Sumaré-SP, Brazil. Rev Bras Enferm 2011; 64(5): 864-9.
6
Vasseur R. La stérilisation, un choix stratégique pour l’hôpital: contribution de l’infirmière générale Mémoire de Santé Publique. Paris: ENSP 1999.
7
Ministère de l'Emploi et de la Solidarité Secrétariat d'Etat à la Santé et à l'action sociale (France) Bonnes pratiques de Pharmacie Hospitalière Première édition, Paris 2013. Available from: http://nosobase.chu-lyon.fr/Reglementation/2001/Rapport/bpph.pdf Paris. [Accessed: 17 /03]
8
Ahoyo TA, Bankolé HS, Adéoti FM, et al. Prevalence of nosocomial infections and anti-infective therapy in Benin: results of the first nationwide survey in 2012. Antimicrob Resist Infect Control 2014; 3: 17.
9
Rutala WA, Weber DJ. guideline for disinfection and sterilization in healthcare facilities, 2008. Centers for Disease Control and Prevention, 2008. Available from: http://www.cdc.gov/hicpac/Disinfection_Sterilization/1_sumIntroMethTerms.html [Accessed: 4/03/2016];
10
Ministère de la santé. Annuaire des statistiques sanitaires: Année 2012. Cotonou. Ministère de la santé, 2013.
11
Donabedian A. Une introduction à l'assurance de la qualité dans les soins de santé. 1st éd. New York, NY: Oxford University Press 2003; 1.
12
Talon D, Reberg A, Blatt A, Hamami k, Sada I, Armaud Ph. Faisablité et fiabilié de la traçabilité des instruments réutilisables á l'aide de puce RFID. Steriprocom international 2008; 1(1): 23-7.
13
Ouendo E-M, Saizonou J, Degbey C, Glèlè Kkai C, Glèlè Y. MAKOUTODE M. Gestion du risque infectieux associé aux soins et services au Centre National Hospitalier et Universitaire Hubert Koutoukou Maga de Cotonou (Bénin). Int J Biol Chem Sci 2015; 9(1): 292-300.
14
Viayinon M. Problématique du traitement du matériel médico-chirurgical dans l’unité de stérilisation du bloc de Gynéco-obstétrique du CNHU de Cotonou (Bénin) Mémoire de fin de formation des instrumentistes du bloc opératoire. Abidjan: Institut National de Formation des Agents de Santé 2003.
15
Ministère de la Santé (France). Circulaire DGS/5 C/DHOS/E 2 n° 2001-138 du 14 mars 2001 relative aux précautions à observer lors de soins en vue de réduire les risques de transmission d'agents transmissibles non conventionnels 2003. Bulletin Officiel n°2001-11 mis à jour le 13 août 2003.
16
Agence de Presse Médicale. Fiche Métier, 2013. Available from: http://www.snitem.fr/sites/default/files/ckeditor/fiche_metier_agent_de_ sterilisation.pdf [Accessed: 15/6]
17
Dégbey C, Aguemon B, Ouendo E-M, Makoutode M, Simon A. Etude de la qualité du matériel médico-technique utilisé dans les blocs opératoires en vue de la prévention des infections associées aux soins et services au Centre National Hospitalier et universitaire de Cotonou au Bénin. J de la Société Clinique de Biologie 2013; 18: 29-35.
18
Richard S. La stérilisation au bloc opératoire: Comment instaurer la confiance. Revue de soins N° 623, éd Masson, Paris 1998; 14-21.
19
Joram N, de Saint Blanquat L, Stamm D, Launay E, Gras-Le Guen C. Healthcare-associated infection prevention in pediatric intensive care units: a review Eur. J Clin Microbiol Infect Dis 2012; 4(3): 56-63.
20
Charles C. Assurance qualité de stérilisation centrale, la sécurité á l'hôpital. Revue de soins N° 623 ed Masson, Paris 1998; 21-7.
21
Fritz H, Erik C, Rolf M, et al. Microbiologie médicale. Flammarion médecines Sciences Paris 2008; 764.